OSCILLOMETRIC NONINVASIVE BLOOD PRESSURE

MEASUREMENTS: THE INFLUENCE OF EXTERNAL ARTIFACTS

V. Jazbinsek, J. Luznik, S. Mieke and Z. Trontelj

(2)

Many oscillometric non-invasive blood pressure (NIBP) measuring devices are based on recording the arterial pressure pulsations in an inflated cuff wrapped around a limb during the cuff pressure deflation [1]. The recorded NIBP data contains the pressure pulses in the cuff, called oscillometric pulses, superimposed on the cuff pressure deflation. Such recordings are often contaminated with external artifacts generated by various patient movements. The objective of this research was to separate the deflation and external artifacts from the pressure pulses and to obtain the typical shapes of pressures changes in the cuff due to those artifacts. The deflation, which is much slower process compared to the arterial pressure pulsation, can be removed from the recorded NIBP data by the digital high pass filtering. Obtained waveform of pressure pulses is known as the oscillometric waveform and automated NIBP measuring devices use its shape to determine systolic and diastolic blood pressure values. On the other hand, pressure changes in the cuff caused by external artifacts, which often generate similar frequency responses as the arterial pressure changes, could not be removed from the oscillometric waveforms by simple filtering. In automated NIBP measuring devices, such artifacts cause either error or false determination of systolic and diastolic pressures. In our analysis, we project the pressure data into the normalized heart beat view. We performed measurements on several healthy volunteers without and with external artifacts (like finger, hand, arm movements, tremor, coughing, etc.) included. We have averaged the transformed waveforms from measurement without external artifacts into normalized reference waveform. In cases, where oscillometric recordings are contaminated with external artifacts, the normalized reference can be used as a template to extract a normal oscillometric waveform, which can then be used for automated determination of systolic and diastolic pressures. Consequently, the typical shapes of external artifacts are also obtained.

[1] Ng K-G, Small CF. Survey of automated non-invasive blood pressure monitors, Journal of Clinical Engineering, 1994; 19:452-475.

Reference

POVEZANI DOKUMENTI

Based on the research results and arguments, as presented in the literature review, the external variables that are also included into our model are (a) pedagogical support in

Abstract: We have applied the Fourier analysis and digital filtering of non-invasive blood pressure data to separate the cuff deflation from the arterial pressure pulses and to

sults show that the SBO generate better reference wave- forms than the WO. With this procedure most of the arte- rial pressure pulses can be subtracted from the measured data. This

To get the reference signal for a given measured oscillation waveform in a real (measured) time scale, one has to transform the normalised reference oscillation

Due to the nationwide School Organisation Law, school social workers are considered as “school external persons” (except in Vienna, where they are subjected directly to the

Relative quantification is the most commonly used quantification method in genetic research (e.g. for gene expression analysis). Ct values are normalized to a reference gene known

The whole frame- work can be used for new applications, the input system was designed to describe all natural and cultural heritage artifacts, the open APIs and standard

The last hypothesis focused on external support as additional to internal support. We tried to confirm that a deeper search in reference sources, understood by us as more